Amperometric Detector

In the case of applied voltage, the amperometric detector detects the change of current caused by the redox reaction of the substance to be measured on the electrode surface. Amperometric detector is often used to analyze ions with low dissociation, which are difficult to detect conductivity detectors and have electrical activity.

  • Unique noise reduction technology
  • Built-in Faraday shielded temperature oven
  • Applicable to IC, HPLC and UHPLC
  • With DC, PULSE and SCAN mode
  • Different flow cells to cover every application
Amperometric Detector

Technical data

Type DC mode PULSE mode
Working Electrode Silver electrode Gold electrode
Noise <2pA <2pA
Signal Range 10pA-200uA in 1.2.5 increments 10nA-200uA in 1.2.5 increments
Potential Range -2.50to +2.50V with 10 mV increments -2.50to +2.50V with 10 mV increments
Analog Output (DAC) 1 to +1V 1 to +1V
Analog Output (I/E) -2.50to +2.50V -2.50to +2.50V
Data Rate 1-100 Hz in 1, 2, 5 increments 1Hz Scan rate 1-100 mV/s in 1, 2, 5 increments
Detectable Substance CN-, I-, Br-, S2- Sugar, aliphatic, amine, aldehyde, amino sugar