Conductivity monitor with flow cell for up to 100 ml/min and optional pH measurement
Conductivity monitor with optional pH measurement
- The AZURA CM 2.1S is a reliable conductivity monitor which is usually utilized in FPLC to follow salt gradients. By adding a pH sensor also pH values can be measured.
- Wetted parts are completely biocompatible. The contactless measurements of conductivity reduces the risk of carryover to a minimum.
- The flow cell covers a flow rate up to 100 ml/min and is delivered with the CM 2.1S.
Flow cell
Flow cell type | Contactless conductivity flow cell |
---|---|
Biocompatible | Yes |
Connection of flow cell | Female 10-32″ UNF or M8x1 thread (PEEK) – both included in shipment |
Capillary connection | 1/16″ or 1/8″ – both included in shipment |
Wetted materials | PEEK |
Flow cell volume | 300 µl |
Max. flow rate | 100 ml/min (recommended: 10 – 100 ml/min) |
Maximum pressure | 100 bar |
Detection
Detector type | Conductivity monitor |
---|---|
Measurement accuracy | Conductivity: ± 2 % full scale value (0.1 – 300 mS/cm), ± 5 % full scale value (300 – 999 mS/cm) Temperature: ± 1.0°C pH 2 – 12: ± 0.2 pH (15 – 25°C), ± 0.5 pH (4 – 40°C) |
Measurement range | Conductivity: 0.1 – 999 mS/cm pH: 0 – 14 |
Linearity | Conductivity: ± 1 % full scale value (0.1 – 300 mS/cm), ± 1 % full scale value (300 – 999 mS/cm) |
pH measurement | Optionally |
Maximum data rate | 5 Hz |
Supported electrodes | All pH electrodes with BNC connector and a voltage output of maximal ± 500 mV |
Communication
Analog outputs | 2 channels (conductivity and pH value – not active if remote controlled by software) DAC 18 bit |
---|---|
Digital outputs | LAN; RS-232 |
Technical parameters
GLP | Electronic serial number |
---|---|
Display | LCD, 2 x 8 characters |
Ambient conditions | Operating temperature: 4 – 40°C, 39,2 – 104 °F Relative humidity: below 90 %, non condensing |
General
Power supply | 100 – 240 V, 50 – 60 Hz, max. 20 W |
---|---|
Dimensions | 121 x 129 x 187 mm (W x H x D) |
Weight | 3.2 kg |